View Single Post
Old 04-09-2008, 09:04 PM   #12 (permalink)
neksnap
 
Join Date: Apr 4 2008
Posts: 8
Rep Power: 0 neksnap is on a distinguished road
Re: manager pro wont pick up bed?

Quote:
Originally Posted by Brian Oshman View Post
Make sure when you address the bed that the period is solid and not flashing. If it is flashing in address mode then the address you are setting is over 100. You could also take the manager to the room and plug only that bed in and press call/menu 888 enter and then hit next, next, next, etc... until you get to auto address room 19 and then hit the start button on the bed to accept the address.

If none of this works then the comm chip in the timer board is bad and it will never communicate until you sawp that chip out.
Thanks for the response.
The address on this bed does not have a period, You go from 1 to 255.
I replaced the com chip in the board and nothing also, I swaped out the entire board with a new one that ETS sent me and nothing. I tried to bring the delay time down to 3 min from 9 min and nothing????
For the last step I have ETS shipping me a new 10 pin harness that plugs in the little board that the three tmax cables plug into in the back of the bed and that harness goes to and plugs into the timer board. Also sending me the Tmax J22 plug in board in the back of the bed.
If this does not work than I as a Tech give the **** up and trust me guys I never give up!
neksnap is offline   Reply With Quote