Re: tmax problem, removed a bed
Sounds like the bed had a built-in timer, and no separate room 3-A was used.
Temporarily, run a cable from room 9 directly to room 11.
Turn off, and then power up the Manager PRO and it will skip #10, as each 3A (or Built-in timer, as it sounds like you have) has the address programmed into it.
when you put the new bed in, make sure it is programmed for the #10 position, re-boot the manager again, and you should be fine.
I believe ADI, and many disty's will have a coupling device that could do the same thing with the two cables left stranded in room #10 also.
|